可插入心脏监测器远程编程的现实世界使用:一个多中心的欧洲经验
Samir Fareh1, Stefano Nardi2,3, Luigi Argenziano2,3
1Hôpital de la Croix-Rousse et Hôpital Lyon Sud, Hospices Civils de Lyon, Lyon, France.
Heart rhythm O2
|December 8, 2025
概括
可插入心脏监测器 (ICM) 的远程重新编程可以显著降低传输负担和警报. 这项技术有可能得到更广泛的采用,以最大限度地减少办公室内访问,并优化患者护理.
科学领域:
- 心脏病学 心脏病学
- 医疗器械 医疗器械
- 医疗信息学 医疗信息学
背景情况:
- 插入式心脏监测器 (ICM) 提供持续的心律失常监测,但产生大量的数据传输量,增加临床工作量.
- LUX-Dx ICM 便于远程重新编程设备设置,旨在减少对办公室内患者访问的需求.
研究的目的:
- 评估ICM远程重编程在欧洲商业化后的现实应用和有效性.
- 评估远程重编程对与ICM相关的传输负担的影响.
主要方法:
- 从23个欧洲中心 (2022-2024) 连续697名患者的非识别数据进行了回顾性分析.
- 评估重编程事件,传播率 (总体,警报,预定,患者/临床医生发起的) 和事件记录频率.
- 重编程干预之前和之后的传输指标的比较.
主要成果:
- 在230个ICM中,共发生了401次重编程事件,其中39%是远程执行的.
- 重编程显著降低了整体传输率 (中位数为57%),警报传输 (78%) 和记录的事件 (91%) (所有P<.001).
- 昏迷是最常见的征兆 (48%);胸是最常见的记录发作类型.
结论:
- 对于优化设备功能和减轻远程监控工作负载,ICM重编程至关重要.
- 在39%的案例中使用远程重新编程,为更广泛的采用提供了一个机会,以减少办公室内访问.
- 通过采用基于警报的监控和逐步取消计划传输,可以实现进一步的效率提升.
更多相关视频
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
12.2K
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
24.7K
相关概念视频
Holter Monitor: 24-Hour Monitoring
1.9K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
1.9K
Pulse rhythm
1.3K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.3K