在环境电离源中开发和优化用于减少背景信号的电子喷射转换器
Seth M Eisenberg1, Alena N Joignant1, Jonathan D Trumbull2
1Biological Imaging Laboratory for Disease and Exposure Research (BILDER), Department of Chemistry, North Carolina State University, Raleigh, North Carolina 27695, United States.
Journal of the American Society for Mass Spectrometry
|December 8, 2025
概括
一种新型的电子喷雾转移器在红外矩阵辅助激光消耗电喷雾电离 (IR-MALDESI) 中显著减少背景离子. 这种离子快门可以提高质谱成像和高通量选的信号质量.
科学领域:
- 分析化学 分析化学
- 质谱测量质量谱测量
- 频谱学是一种光谱学.
背景情况:
- 脉冲环境电离源经常受到背景离子的影响.
- 背景离子可以抑制样本信号并阻碍统计分析.
- 红外矩阵辅助激光吸附电喷电离 (IR-MALDESI) 易受背景噪声的影响,影响质谱成像和高通量选.
研究的目的:
- 开发和评估一个用于IR-MALDESI.I的电喷转向器.
- 在IR-MALDESI实验中减少背景离子信号.
- 增强与样本相关的信号,提高质谱应用中的数据质量.
主要方法:
- 开发了一种作为离子快门的电子喷射转向器.
- 调节器的时间优化用于基于液体的查和组织成像.
- 对背景信号减弱和样本信号增强的评估.
主要成果:
- 背景信号减少了多达4倍.
- 与样本相关的信号增加了多达11倍.
- 观察到样本信号的相对持续增加,没有对数据产生负面影响.
结论:
- 电子喷雾转向器有效地减少了IR-MALDESI中的背景离子.
- 这种技术可以提高敏感分析应用的信号质量.
- 开发的离子快门可以提高IR-MALDESI在各种查和成像任务中的性能.
相关概念视频
Electrospray Ionization (ESI) Mass Spectrometry
2.0K
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
2.0K
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
1.7K
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
1.7K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
1.6K
Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
1.6K
Atomic Emission Spectroscopy: Interference
579
In atomic emission spectroscopy (AES), high-temperature atomizers excite a broad range of elements and molecules that generate complex emissions from sources such as oxides, hydroxides, and flame combustion products in the flame or plasma. Several strategies can be employed to minimize spectral interferences caused by overlapping emission lines or bands. These include increasing instrument resolution, choosing alternative emission lines, optimally placing the detector in low-background regions,...
579
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
1.2K
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
1.2K
Atomic Absorption Spectroscopy: Interference
2.0K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
2.0K