在1064nm的多通路差异光声谱仪用于环境气溶吸收
Jie Chen1, Xianmei Qian1, Wenyue Zhu1
1State Key Laboratory of Laser Interaction with Matter, Anhui Institute of Optics and Fine Mechanics, HFIPS, Chinese Academy of Sciences, Hefei 230031, China.
Photoacoustics
|December 31, 2025
概括
一个新的多通差光声谱仪 (MP-DPAS) 精确测量大气气溶吸收. 这种进步改善了气候模型和对区域天气模式的理解.
科学领域:
- 大气科学 大气科学
- 频谱学是一种光谱学.
- 气候科学 气候科学
背景情况:
- 精确的气体气溶吸收测量对于了解气候影响和区域气象过程至关重要.
- 现有的方法可能缺乏详细大气研究所需的灵敏度或精度.
研究的目的:
- 开发和验证一种新型的多通差光声谱仪 (MP-DPAS),用于在现场测量大气气溶吸收系数 (Abs).
主要方法:
- 这项研究使用了在1064nm工作的多通路差分光声谱仪 (MP-DPAS).
- 该MP-DPAS设计包括22个激光反射来增强光声信号和差异配置来抑制噪声和放大信号.
- 验证涉及比较测量Abs与Mie散射理论计算和扫描移动粒子测量仪 (SMPS) 数据.
主要成果:
- 在110秒内,MP-DPAS实现了0.05mm-1的最小检测极限,精度为1.4mm-1.
- 多通道配置将光声信号增强了十倍,而差异设置则将其放大了两倍.
- 验证证实了MP-DPAS测量的准确性.
结论:
- 开发的MP-DPAS提供了一种高度敏感和准确的方法,用于在现场测量大气气溶吸收.
- 这种仪器的进步可以大大有助于改善气候建模和气象研究.
- 验证的准确性支持其在关键大气研究中的应用.
更多相关视频
相关概念视频
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
610
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
610
Atomic Absorption Spectroscopy: Instrumentation
1.5K
An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
The atomizer used in AAS can be either a flame atomizer or an...
1.5K
UV–Vis Spectrometers
3.3K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
3.3K
Atomic Emission Spectroscopy: Instrumentation
1.1K
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-phase atoms and ions. There are two main types of atomization devices: continuous and discrete atomizers. Continuous atomizers, like plasmas and flames, introduce samples in a constant stream, while discrete atomizers inject individual samples using syringes or autosamplers. The most common discrete atomizer is the electrothermal atomizer.
1.1K
Atomic Emission Spectroscopy: Overview
3.4K
Atomic emission spectroscopy (AES) is an analytical technique used to determine the elemental composition of a sample by analyzing the light emitted from excited atoms. In AES, atoms in a sample are excited to higher energy levels by thermal energy from high-temperature sources, such as plasma, arcs, or sparks. When these excited atoms return to lower energy states, they emit light at specific wavelengths characteristic of each element. The resulting atomic emission spectrum, which consists of...
3.4K
Atomic Absorption Spectroscopy: Overview
3.2K
Atomic absorption spectroscopy (AAS) is a technique used to analyze elements by measuring electromagnetic radiation (EMR) absorbed by atoms, which causes them to transition to a higher-energy orbit. The most crucial step in AAS is atomization, where the analyte is converted into gas-phase atoms, typically through a flame or furnace. Some of these atoms become thermally excited in the flame, while most remain in the ground state.
When irradiated by EMR of a particular wavelength, these...
When irradiated by EMR of a particular wavelength, these...
3.2K