对生物电阻分析装置的评估,以测量体阻测量的数据可靠性
Dan Bee Kim1, Seong Su Shin1, Wan-Seop Kim1
1Korea Research Institute of Standards and Science, Daejeon, Republic of Korea.
Frontiers in nutrition
|January 29, 2026
概括
使用身体阻抗模拟器 (BII) 校准生物电阻分析 (BIA) 设备可以提高测量可靠性. 这确保了用于临床和研究应用的准确人体成分数据.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
- 计量学 计量学 计量学
背景情况:
- 生物电阻分析 (BIA) 设备对于身体组成分析至关重要,但其准确性可变.
- 设备校准对于在临床和研究环境中可靠的BIA测量至关重要.
- 在BIA设备准确度的变化需要标准化的校准方法.
研究的目的:
- 为了评估多频BIA设备的阻抗测量能力.
- 用自制的体阻抗模拟器 (BII) 来评估BIA设备的准确性.
- 通过对认可的参考标准进行校准来量化测量不确定性.
主要方法:
- 使用可追溯到国家阻抗标准的LCR计校准BII.
- 通过使用校准的BII.II,评估了两个多频BIA设备模型.
- 在广泛的频率范围 (1kHz至1MHz) 中测量阻抗,以评估精度和线性.
主要成果:
- 在BIA设备模型之间观察到显著的性能差异,尽管规格相似.
- 模型A表现出更高的准确性和一致性,不确定性约为1.4%.
- 模型B显示出更大的偏差,特别是在干阻抗测量中.
结论:
- 通过使用标准化程序,可靠地校准BIA设备,并使用像BII这样的稳定参考.
- 校准显著提高了BIA测量数据的可靠性和信心.
- 标准化的校准程序对于确保BIA数据的质量至关重要.
更多相关视频
07:44Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
1.7K
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
999
相关概念视频
Impedance Combination
751
Consider a string of christmas lights, each bulb symbolizing an impedance element. In this series configuration, the flow of electric current remains uniform across every component. This behavior aligns with Kirchhoff's Voltage Law (KVL), which asserts that the total impedance in such a setup equals the sum of individual impedances—akin to resistors in series. It follows that the voltage from the power source is distributed proportionally among these components, adhering to the voltage...
751
Impedances and Admittance
1.9K
In the realm of AC circuits, passive circuit elements like resistors, inductors, and capacitors take on a different character when characterized by phasor voltage and current. Their behavior is expressed through impedance, a vital concept in AC circuit analysis.
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
1.9K
Series Impedances: Three-Phase Line
444
Calculating series impedances for a three-phase overhead line involves evaluating resistances and inductive reactances in a network with three-phase and multiple neutral conductors grounded at regular intervals.
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
444
Bus Impedance Matrix
527
Calculating subtransient fault currents for three-phase faults in an N-bus power system involves using the positive-sequence network. When a three-phase short circuit occurs at a specific bus, the analysis uses the superposition method to evaluate two separate circuits.
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...
527
Line Protection with Impedance Relays
451
Coordinating time-delay overcurrent relays in complex radial systems and directional overcurrent relays in multi-source transmission loops can be challenging. Impedance relays address these issues by responding to the voltage-to-current ratio, specifically measuring the apparent impedance of a line. These relays become more sensitive during faults as current increases and voltage decreases, thereby reducing the apparent impedance.
Under normal conditions, low load currents keep the measured...
Under normal conditions, low load currents keep the measured...
451
RLC Series Circuits: Impedance
2.6K
When current flow is opposed in a DC or AC circuit, it is referred to as resistance or impedance, respectively. Impedance plays a key role in determining the performance of AC circuits. It is represented by Z, which is a combination of resistance and reactance, and depends upon the angular frequency, measured in ohms.
Thus, the magnitude of the impedance is given by the following equation,
Thus, the magnitude of the impedance is given by the following equation,
2.6K