对于宽带水下声波雷达的厚度不敏感的编码元表面,用于降低横截面的水下声波雷达
Jia-Wang Zhang1, Sheng-Dong Zhao2, Hao-Wen Dong3
1School of Mathematics and Statistics, Qingdao University, Qingdao 266071, People's Republic of China.
The Journal of the Acoustical Society of America
|February 17, 2026
概括
这项研究引入了一种新的1位编码元表面 (CM) 宽带水下声学隐身. 通过扩散声波,CM有效地抑制了广泛的频率范围内的目标检测能力,克服了传统的厚度限制.
科学领域:
- 水下声学 水下声学
- 超材料科学 超材料科学
- 声学工程 声学工程
背景情况:
- 传统的声学隐形依赖于对特定频率调整的材料厚度,限制宽带性能.
- 实现声波散射抑制需要克服固有的厚度-频率依赖.
研究的目的:
- 开发一个宽带声学元表面,将厚度与频率脱,以提高隐形性.
- 为了证明在广泛的频率范围 (10-35 kHz) 上有效地抑制声波散射.
主要方法:
- 引入一个1位编码的超表面 (CM) 与两个不同的深次波长单位.
- "0"和"1"单元的设计,分别模仿刚性和释放压力的边界.
- 空间编码的优化,以将事件波重定向为分散的散射模式.
主要成果:
- 从10到35kHz实现了有效的声学散射抑制,对腔体几何形状的灵敏度最小.
- 在一个广的角跨度 (高达45°) 上,已经证明了超过10dB的雷达横截面减小.
- 确认稳定的反射相对空气腔厚度的变化,展示了一个厚度脱的设计.
结论:
- 开发的CM为宽带水下声学签名控制提供了一个强大的,易于制造的战略.
- 这种厚度脱的设计范式为声纳伪装和波面操纵开辟了新的途径.
- 这项研究在实现各种音响环境中的隐形性方面取得了重大进展.
相关概念视频
Equipments Used To Measure Blood Pressure
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Difference from Background: Limit of Detection
The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
The LOD indicates the presence or absence...
Parallel Resonance
The parallel RLC circuit is an arrangement where the resistor (R), inductor (L), and capacitor (C) are all connected to the same nodes and, as a result, share the same voltage across them. The parallel RLC circuit is analyzed in terms of admittance (Y), which reflects the ease with which current can flow. The admittance is given by:
Design Example
The innovation of touch-tone telephony revolutionized the telecommunications industry by replacing the traditional rotary dial with a dual-tone multi-frequency (DTMF) signaling system. This system uses a matrix-style keypad with buttons arranged in four rows and three columns, creating 12 distinct signals each assigned to a pair of frequencies. Each button press results in a simultaneous generation of two sinusoidal tones – one from a low-frequency group (697 to 941 Hz) and one from a...
Downsampling
When considering a sampled sequence with zero values between sampling instants, one can replace it by taking every N-th value of the sequence. At these integer multiples of N, the original and sampled sequences coincide. This process, known as decimation, involves extracting every N-th sample from a sequence, thereby creating a more efficient sequence.
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
Electronic Distance Measuring Instruments
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over short distances...


