一种基于同位素稀释-液体色谱-并联质谱的候选参考测量程序,用于测量人体血清和血中的量化
Elie Fux1, Rupert Schmid1, Neeraj Singh1
1Roche Diagnostics GmbH, Penzberg, Germany.
Clinical chemistry and laboratory medicine
|February 18, 2026
概括
一种新的参考测量程序 (RMP) 准确地使用同位素稀释-液态色谱-并联质谱法 (ID-LC-MS/MS) 量化血清和血中的. 这种方法确保了临床诊断和测试标准化的可靠结果.
科学领域:
- 临床化学 临床化学
- 分析化学 分析化学
- 质谱测量质量谱测量
背景情况:
- 准确的丸激素量化对于诊断各种疾病至关重要.
- 现有的方法可能缺乏可靠标准化所需的精度和可追溯性.
研究的目的:
- 开发和验证一种用于测量的候选参考测量程序 (RMP).
- 通过使用初级参考材料,确保SI单位的可追溯性.
- 建立一种可靠的人类血清和血分析方法.
主要方法:
- 同位素稀释-液体色谱-并联质谱法 (ID-LC-MS/MS) 采用二维心脏切割方法.
- 固体相提取 (SPE) 用于样品清理.
- 根据当前指南进行验证,包括选择性,矩阵效应,精度,准确性和不确定性评估.
主要成果:
- 量化范围:27.7 pmol/L到62.4 nmol/L,没有干扰.
- 高精度,中等精度≤3.1%,可重复性为1.4-1.9%.
- 低偏差 (-1.2%至3.0%) 和测量不确定性 (≤2.9%扩展到目标值).
结论:
- 开发的RMP表现出对的高分析性能.
- 该方法适用于常规测试标准化.
- 它为临床样本评估提供准确和精确的结果.
更多相关视频
12:47Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
10.0K
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
13.5K
相关概念视频
Peptide Identification Using Tandem Mass Spectrometry
6.2K
Tandem mass spectrometry, also known as MS/MS or MS2, is an analytical technique that employs two mass analyzers. Essentially it is a series of mass spectrometers that helps isolate a particular biomolecule and then helps study its chemical properties.
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
6.2K
Tandem Mass Spectrometry
3.0K
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and reduce chemical noise during analyte detection. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called...
3.0K