对眼内压力测量的比较分析,跨越不同度法和经验水平
Şerife Çiloğlu Hayat1, Yusuf Cem Yılmaz1, Hatice Davarcı2
1Specialist Dr, Basakşehir Cam Sakura State Hospital, Department of Ophthalmology, Istanbul, Turkey.
Korean journal of ophthalmology : KJO
|February 23, 2026
概括
戈德曼平面度测量 (GAT) 在眼科医生中显示出显著的观察者间变异性,经验较少的临床医生影响了测量准确性. 非接触性度测量 (NCT) 也显示出变异性,这表明在临床解释方面要谨慎,需要进行结构化培训.
科学领域:
- 眼科医生 眼科 眼科
- 视光学测量 视光学测量 视光学测量
- 临床测量 临床测量
背景情况:
- 眼内压力 (IOP) 测量对于青光眼的诊断和管理至关重要.
- 戈德曼平面度测量 (GAT) 是一种标准方法,但其可重现性可能会受到观察者经验的影响.
- 非接触性度测量 (NCT) 提供了一种更快,更替代的方法,但其与GAT的协议需要评估.
研究的目的:
- 评估不同经验水平的眼科医生之间戈德曼平面化度测量 (GAT) 的观察者间变异性.
- 评估测量眼内压力 (IOP) 的非接触性度计 (NCT) 和GAT之间的协议.
主要方法:
- 一项涉及210名患者的横截面观察性研究.
- 使用NCT对右眼进行IOP测量,然后在右眼进行GAT测量.
- 三名观察员参与:一名高级眼科医生 (Obs1) 和两名住院医生 (Obs2,Obs3).
主要成果:
- 在高级眼科医生和一名住院医生 (Obs2) 之间,以及高级眼科医生和NCT之间,发现了GAT-IOP测量的显著差异.
- 平均差异 (MD) 和95%的一致度极限 (LoA) 不同:Obs1与Obs2 (0.8 mmHg,±5.7 mmHg),Obs1与Obs3 (0.4 mmHg,±6.5 mmHg),Obs1与NCT (-0.9 mmHg,±7.3 mmHg).
- 与Obs3和NCT相比,Obs2读数的更高百分比在Obs1的±4 mmHg范围内;NCT显示了广泛的LoA,特别是在正常压力小组中.
结论:
- 在GAT测量中观察者间的变异性超过了±4 mmHg的临床基准,这表明在经验较少的观察者中可能存在与技能相关的准确性问题.
- 临床医生在解释NCT结果时应谨慎,特别是在术后或治疗调整期间.
- 结构化的培训协议和与GAT的交叉检查可以提高IOP测量的一致性.
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
1.8K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.8K
Measurement of Blood Pressure
3.5K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
3.5K
Special considerations while measuring blood pressure
1.3K
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
1.3K
Equipments Used To Measure Blood Pressure
3.7K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.7K
Glaucoma: Overview
1.5K
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
1.5K
Sites for measuring blood pressure
3.8K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.8K