基于永久磁铁金属片的非接触式流量传感器的开发,用于监测循环状态
Kicheol Yoon1,2,3,4, Seung Hee Choi3, Tae-Hyeon Lee3
1Gachon Biomedical Convergence Institute, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea.
Biosensors
|February 26, 2026
概括
这项研究引入了一种新型传感器来监测状膜炎. 非侵入性设备使用洛伦茨力实时检测流体流动,提高患者的安全性和降低成本.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
- 传感器技术 传感器技术
背景情况:
- Ascites 膜手术的过程可能会因导管上的异物而复杂化.
- 目前的监测依赖于视觉检查,这可能是主观的,耗时的.
- 需要实时自动状态检测,以提高程序安全性和效率.
研究的目的:
- 开发和评估一种新型的传感器,用于实时监测关节孔状况在状垂体造过程中.
- 能够自动检测液体流动,减少对视觉检查的依赖.
- 创建一个非侵入性,可重复使用的传感器,以提高患者的安全性和降低医疗成本.
主要方法:
- 设计了一种传感器,集成永久磁铁,和信号处理电路.
- 使用Mu-金属屏蔽来防止电磁干扰.
- 该传感器使用生理盐溶液进行了测试,其性能通过有限元分析进行了评估.
主要成果:
- 传感器使用洛伦茨力成功检测到液体流动.
- 当流体存在时,平均电压为11.07μV,这表明流体检测成功.
- 有限元分析证实了传感器对实时状态监控的能力.
结论:
- 拟议的基于劳伦茨力的传感器提供了一种非侵入性和有效的方法来监测腹膜穿刺.
- 片设计允许轻松安装,脱离和重复使用,提供成本效益.
- 这项技术有可能提高癌症患者的形形切割的安全性和效率.
相关概念视频
Equipments Used To Measure Blood Pressure
3.8K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.8K
Magnetic Damping
1.2K
Eddy currents can produce significant drag on motion, called magnetic damping. For instance, when a metallic pendulum bob swings between the poles of a strong magnet, significant drag acts on the bob as it enters and leaves the field, quickly damping the motion.
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
1.2K
Eddy Currents
2.8K
Since eddy currents occur only in conductors, magnets can separate metals from other materials. For example, in a recycling center, trash is dumped in batches down a ramp, beneath which lies a powerful magnet. Conductors in the trash are slowed by eddy currents, while nonmetals in the trash move on, separating from the metals. This works for all metals, not just ferromagnetic ones.
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
2.8K
Amperometry: Overview
1.9K
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
1.9K