使用植入式血液动力学监测器的初始经验
D M Steinhaus1, R Lemery, D R Bresnahan
1Department of Cardiology, University of Missouri-Kansas City School of Medicine, USA.
Circulation
|February 15, 1996
概括
可植入的血液动力学监测器可以测量长期的右心室氧和和肺动脉压力在门诊患者. 这项技术显示出可行性和通常可靠的传感器功能,用于长期的血液动力学监测.
科学领域:
- 生物医学工程 生物医学工程
- 心血管生理学心血管生理学
- 可以植入的医疗器械
背景情况:
- 心内血动力学监测通常仅限于急性护理环境中的短时间.
- 慢性监测对于管理严重的左心室功能障碍等疾病至关重要.
- 现有的方法缺乏连续的,长期的门诊血液动力学数据.
研究的目的:
- 开发和评估一种可植入的血液动力学监测仪.
- 评估慢性测量右心室氧和和肺动脉压力的可行性.
- 为了确定植入器械的安全性和可靠性.
主要方法:
- 开发了一种可植入装置,具有皮下控制器和右心室和肺动脉的透静导管.
- 在10名患有严重左心室功能障碍的患者中植入该设备.
- 通过挑动作和长期随访,同时进行导管测量来评估设备的性能.
主要成果:
- 该装置在植入器时成功测量了肺动脉压力 (平均40 +/- 12 mm Hg) 和右心室氧和度 (平均51%) .
- 证明了对运动和酸甘等机动的预期生理反应.
- 在随访期间 (0.5-15.5个月) 与气球漂浮导管相比没有显著差异.
- 经验丰富的脱落 (4/10),传感器漂移 (3/10) 和传感器故障 (2/10).
- 4名患者死亡,4名患者接受了移植;没有观察到显著的组织损伤.
结论:
- 使用可植入传感器技术的慢性血液动力学监测在门诊患者中是可行的.
- 该设备通常耐受性良好,没有显著的不良影响.
- 传感器随着时间的推移提供可靠的数据,尽管长期的临床实用性需要进一步建立.
更多相关视频
11:17Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
12.3K
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
27.0K
相关概念视频
Holter Monitor: 24-Hour Monitoring
4.1K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
4.1K
Pulse rhythm
1.6K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.6K