Puerperal Weck clip sterilization: study I (first study of two consecutive studies).
Fertility and Sterility
|December 1, 1976
Summary
The original Weck pistol-grip applicator for puerperal sterilization showed an 8.7% failure rate. Mechanical issues were noted, but tubal tissue histology suggests potential reversibility of clip application.
Area of Science:
- Gynecology
- Surgical Devices
- Reproductive Medicine
Background:
- Puerperal sterilization is a common method for permanent contraception.
- The Weck pistol-grip applicator was used for clip application during puerperal sterilization.
- Evaluating the efficacy and mechanical performance of surgical devices is crucial.
Purpose of the Study:
- To assess the corrected failure rate of the original Weck pistol-grip applicator for puerperal sterilization.
- To document mechanical difficulties encountered during application.
- To analyze the configuration of recovered clips and the resulting tubal tissue histology.
Main Methods:
- One hundred and four puerperal applications using the original Weck pistol-grip applicator and large Weck clip in an inverted V configuration.
- Follow-up ranging from 37 to 42 months.
- Analysis of recovered clip configuration and tubal tissue histology.
Main Results:
- The corrected failure rate for this sterilization method was 8.7%.
- Mechanical difficulties were observed with the applicator's jaws and metal sleeve.
- Histologic examination of tubal tissue indicated potential reversibility.
Conclusions:
- The original Weck pistol-grip applicator demonstrated a significant failure rate in puerperal sterilization.
- Mechanical deficiencies in the applicator device were identified.
- Histological findings suggest that the effects of clip application on fallopian tubes may not be permanent.