Related Experiment Video
Updated: Jul 31, 2026

Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
Influence of the reference gas of paramagnetic oxygen analyzers on nitrogen concentrations during closed-circuit
J F Hendrickx1, A A van Zundert, A M de Wolf
1Department of Anesthesiology, Intensive Care and Pain Therapy, Catharina Hospital, Eindhoven, The Netherlands.
Abstract:
Nitrogen (N2) may accumulate to unacceptable levels during closed-circuit anesthesia (CCA) when the sampled gases are redirected to the anesthesia circuit, because many gas analyzers entrain air as a reference gas to calibrate for oxygen analysis. Using oxygen instead of air as the reference gas for paramagnetic oxygen analysis could attenuate N2 accumulation. Forty-three adult ASA physical status I-III patients undergoing a variety of peripheral and abdominal procedures were assigned to one of two groups, depending on the reference gas used by a paramagnetic oxygen analyzer, either air (group I, n = 23) or oxygen (group II, n = 20). Gases sampled by the multigas analyzer were redirected to the anesthesia circuit. End-expired N2 (N2Et) was calculated as "balance gas": (100 - %O2 - %anesthetic vapor - %CO2). N2Et after 55 min (N2Et55min) was correlated with the end-expired N2 concentration when the circuit was closed (N2Et0min) and 5 min (N2Et5min) thereafter. In group I, N2Et accumulated almost linearly over time (t, min): N2Et (%) = 2.47 + 0.61 * t (r2 = 0.999). N2Et0min, N2Et5min, and N2Et55min were 1.3+/-0.8, 5.3+/-1.7, and 35.3+/-5.3%, respectively (mean +/- SD). The correlation (r2) between N2Et55min and N2Et0min was 0.19, and between N2Et55min and N2Et5min it was 0.56. In group II, N2Et increased exponentially: N2Et (%) = 1.01 + 11.9 * (1 - e(-t/43.5)) (r2 = 0.99). N2Et0min, N2Et5min, and N2Et55min were 0.87+/-0.93, 2.6+/-1.5, and 10.1+/-2.9%, respectively. The correlation (r2) between N2Et55min and N2Et0min was 0.04, and between N2Et55min and N2Et5min it was 0.40. We conclude that paramagnetic oxygen analyzers that use oxygen as the reference gas significantly attenuate N2 accumulation during CCA, which may reduce the need for frequent flushing of the anesthesia system, may provide more constant oxygen and nitrous oxide concentrations, and may simplify pharmacokinetic studies of potent inhaled anesthetics.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
08:23Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Inhalational Anesthetics: Overview
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated. Under...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...