Lactate determination in exercise testing using an electrochemical analyser: with or without blood lysis?
A G Thin1, Z Hamzah, M X FitzGerald
1Department of Physiology, University College Dublin, Ireland.
European Journal of Applied Physiology and Occupational Physiology
|February 24, 1999
Summary
Electrochemical lactate analyzers are reliable for determining lactate thresholds during exercise testing, even without lysing blood samples. This method provides comparable results to traditional spectrophotometry, simplifying exercise physiology assessments.
Area of Science:
- Exercise Physiology
- Biomedical Engineering
- Analytical Chemistry
Background:
- Lactate measurement is crucial in exercise testing.
- Discrepancies exist between spectrophotometric and electrochemical lactate assays.
- Practical application of electrochemical lactate analyzers requires validation.
Purpose of the Study:
- Compare spectrophotometric and electrochemical lactate measurement methods.
- Evaluate lactate thresholds derived from lysed and non-lysed blood samples.
- Assess the utility of electrochemical analyzers in exercise testing.
Main Methods:
- Bland-Altman statistical analysis was employed.
- Progressive exercise tests to exhaustion were conducted.
- Lactate concentrations were measured using spectrophotometry and electrochemistry on lysed and non-lysed samples.
- Lactate thresholds were derived using a log-log transform technique.
Main Results:
- Spectrophotometric and electrochemical methods showed close agreement for lysed blood lactate.
- Electrochemical measurements of non-lysed blood yielded systematically lower lactate values.
- Lactate thresholds derived from lysed spectrophotometric and electrochemical measurements were not significantly different.
- Lactate thresholds derived from lysed spectrophotometric and non-lysed electrochemical measurements were also not significantly different.
Conclusions:
- Electrochemical lactate analyzers are suitable for lactate threshold determination.
- Sample lysis is not necessary when using electrochemical analyzers with the log-log transform technique.
- This validates a simpler approach for exercise lactate threshold assessment.