Venous duplex scanning of the leg: range, variability and reproducibility
J H Haenen1, H van Langen, M C Janssen
1Clinical Vascular Laboratory, University Hospital Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands.
Clinical Science (London, England : 1979)
|February 25, 1999
Summary
This study establishes normal venous duplex ultrasound values for diameter, compressibility, flow, and reflux in healthy individuals. Results provide essential benchmarks for diagnosing venous diseases and demonstrate good inter-observer reproducibility for key measurements.
Area of Science:
- Vascular Ultrasound
- Venous Hemodynamics
- Medical Imaging
Background:
- Limited data exists on normal venous duplex ultrasound values, variability, and reproducibility.
- Establishing these parameters is crucial for accurate diagnosis of venous disorders.
Purpose of the Study:
- To measure the range and variability of venous diameter, compressibility, flow, and reflux in healthy individuals.
- To determine the reproducibility of these venous measurements.
Main Methods:
- Duplex ultrasound scans were performed on 42 healthy individuals (42 limbs, 714 vein segments).
- Intra-observer variability was assessed in 11 healthy individuals (187 segments).
- Inter-observer variability was evaluated in 15 healthy individuals (255 segments) and 13 deep venous thrombosis patients (169 segments).
Main Results:
- 99% of normal vein segments were traceable; 95% were compressible, and 98% showed flow.
- Proximal vein reflux duration was <1.0s (except common femoral vein); distal reflux was <0.5s (except distal long saphenous vein).
- Inter-observer agreement for compressibility (kappa=0.77) and reflux (kappa=0.86) in patients was good.
Conclusions:
- Normal venous duplex ultrasound values for diameter, compressibility, flow, and reflux have been established.
- Almost all veins are compressible in healthy individuals, with specific reflux time limits for proximal and distal segments.
- The study confirms good inter-observer reproducibility for key venous measurements, vital for clinical application.