Related Experiment Videos
Volume flow measurement in hemodialysis shunts using time-domain correlation
F Basseau1, N Grenier, H Trillaud
1Service de Radiologie, Groupe Hospitalier Pellegrin, Bordeaux, France.
Abstract:
Volume flow was measured in 58 hemodialysis shunts (32 grafts and 26 radial fistulas) using the color velocity imaging-quantification method. This method is based on time-domain correlation for velocity calculation and integration of time-varying velocity profiles generated by M-mode sampling. Measurements were made in the brachial artery to estimate radial fistula flow or directly in the grafts. Intraoperator reproducibility was 14.9% for fistulas and 11.6% for grafts. Flow rate was significantly lower in abnormal shunts associated with a functional disorder or a morphologic complication (808 ml/min +/- 484) than in shunts associated with no abnormalities (1401 ml/min +/- 562). Receiver operating characteristic curves showed that a flow rate of 900 ml/min for fistulas and 1300 ml/min for grafts provided 81% and 79% sensitivity and 79% and 67% specificity, respectively. A functional disorder or a morphologic complication was associated with all fistulas and grafts in which flow rates were lower than 500 ml/min and 800 ml/min, respectively.