Systematic errors in data evaluation due to ethanol stripping and water vaporization
Related Experiment Video
Updated: Oct 24, 2026

Microdialysis of Ethanol During Operant Ethanol Self-administration and Ethanol Determination by Gas Chromatography
Published on: September 5, 2012
1Institute of Chemical Engineering and Biotechnology, Swiss Federal Institute of Technology (EPFL), CH-1015 Lausanne, Switzerland.
Abstract:
Systematic errors due to the neglect of water and/or ethanol partition between liquid and gaseous phases are discussed for bioreactors equipped with or without a condenser. Both water vapor and ethanol vapor are present in the off-gas leaving the condenser. Presence of residual water vapor largely influences the gas measurements by dilution. As a consequence, the oxygen consumption rate can be overestimated by a factor of 3 if calculations are not corrected for water vapor content or if no additional device is implemented after the condenser to completely dry the off-gases. The mass balance and partition equations predict that the condenser has only a small effect on reduction of the ethanol vapor content of the off-gas. The reason is the high ethanol concentration of the condensate droplets on the condenser wall in contact with the off-gases. Model predictions as well as experimental results show that ethanol evaporation represents a large fraction of the ethanol production rate and influences greatly the elemental recoveries. For a reactor working at 30 degrees C without condensation of the vapors and for a volumetric aeration rate of 0.63vvm, stripping of ethanol resulted in a gaseous dilution rate of 0.016 h-1 for ethanol. The dilution rate by stripping was reduced to 0.014 h-1 when a condenser at 12 degrees C was implemented. The fraction of ethanol that is stripped is mainly dependent on the ratio D/vvm (liquid to gaseous flow rates), and the effect is only slightly influenced by low condenser temperature. The evaporation of ethanol may account for more than 20% of the ethanol formation rate. Therefore, the condenser does not succeed to reflux all ethanol to the reactor broth. In terms of a unit operation, ethanol vapor can be efficiently reduced by absorption instead of condensation. To demonstrate the feasibility, a simple modification of the reactor was tested for continuous cultures: the feed port was changed from the top-plate to the top of the condenser, which was used as an absorption column. Ethanol stripping was reduced by a factor of 4 as compared to the condensation setup (at 12 degrees C): it accounted for 2% of the ethanol production rate as compared to 8.2% at D = 0.19 h-1 and 0.63vvm.
More Related Videos
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Data Validation
Key parameters for method validation include: