Variations in tidal volume with portable transport ventilators
Related Experiment Video
Updated: Oct 23, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
E K McGough1, M J Banner, R J Melker
1University of Florida College of Medicine, Gainesville.
Background:
As intra- and interhospital transportation of ventilator-dependent patients has become more commonplace, the number of portable transport ventilators has increased. Transport ventilators should be capable of delivering consistent tidal volume (VT) from breath to breath following changes in lung-thorax compliance and airways resistance. We sought to determine the effect of changes in compliance (C) and resistance (R) on the VT delivered by eight commercially available, time-cycled transport ventilators.
Methods & Materials:
Each ventilator (PneuPAC Model 2, Autovent 3000, MAX, Bird Transport Mini-TXP, IC-2A, P7, E100i, and Logic 07a) was connected to a calibrated pneumotachograph and a test lung set for normal adult C (C = 100 mL/cm H2O [1.02 L/kPa]) and R (R = 2 cm H2O.s.L-1 [0.2 kPa.s.L-1]), with VT at 1,000 mL.
Results:
As C and R were manipulated, VT varied widely. Tidal volume decreased least with the P7 and most with the Bird transport ventilator.
Conclusion:
Decreases in VT with a transport ventilator predispose patients to hypoventilation, hypercapnia, and acidemia. Tidal volume often is not monitored continuously during transport, yet large decreases in VT must not be allowed when pulmonary mechanics are unstable. Internal pressure-limiting valves, venturi flow-generating devices, and compression volume in the breathing circuit are at least three factors that affect VT with transport ventilators.
More Related Videos
04:16Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Respiratory Volumes and Capacities
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...