A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
An under-occluded roller pump is less hemolytic than a centrifugal pump
D J Rawn1, H K Harris, J B Riley
1Clinical Services Department, College of Health Professions, Medical University of South Carolina, Charleston, USA.
Altering roller pump occlusion settings can significantly reduce hemolysis compared to centrifugal pumps. Opening roller pump occlusion settings results in lower index of hemolysis, offering a potential improvement in blood handling.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Hemodynamics
Background:
- Hemolysis, the destruction of red blood cells, is a critical concern in cardiopulmonary bypass.
- Roller pumps (RP) and centrifugal pumps (CP) are commonly used, but their impact on hemolysis varies.
- Optimizing pump settings is essential to minimize blood trauma.
Purpose of the Study:
- To quantify and compare the hemolysis generated by roller pumps at various occlusion levels against a centrifugal pump.
- To test the hypothesis that different roller pump occlusion settings and centrifugal pumps produce similar levels of hemolysis.
Main Methods:
- Five identical closed-loop circuits were established, each with a different pump configuration: a centrifugal pump and a roller pump at four distinct occlusion settings.
- Circuits were primed with bovine blood, and pumps operated at a consistent flow rate (4.5 L/min) and afterload (250 mmHg) for three hours.
- Index of Hemolysis (IH) was measured and compared across all configurations using ANOVA.
Main Results:
- Under-occluded roller pump settings produced significantly lower Index of Hemolysis (IH) compared to the centrifugal pump.
- The centrifugal pump did not show a significantly greater IH than the roller pump set to a barely non-occlusive setting.
- Adjusting roller pump occlusion settings appears to be a viable method for reducing blood damage.
Conclusions:
- Modifying roller pump occlusion settings, particularly reducing occlusion, can lead to a lower Index of Hemolysis (IH) than traditional settings or centrifugal pumping.
- This suggests that roller pumps, when adjusted appropriately, may offer advantages in minimizing blood trauma during extracorporeal circulation.
- Further investigation into optimal roller pump parameters is warranted to enhance hemocompatibility.