Corrosion in stainless-steel and nickel-titanium files
O W Stokes1, P M Fiore, J T Barss
1Department of Endodontics, Northwestern University Dental School, Chicago, IL 60611, USA.
Abstract:
This study evaluated and compared the corrosion susceptibility of stainless-steel and nickel-titanium (NiTi) endodontic files immersed in sodium hypochlorite. For each of the stainless-steel files (Kerr K-Flex, Caulk Flex-O, and Union Broach Flex-R) plus the NiTi files (Union Broach NiTi and Tulsa NiTi), the cutting flutes of 24 ISO size 20 files were immersed into 5.25% sodium hypochlorite. Their open circuit potential (OCP) was recorded for 1 h on a strip chart with high impedance. The strip chart recording for each file was classified into a stability score: (i) stable, (ii) unstable, or (iii) erratic. The OCP was measured by a potentiostat and a standard calomel electrode reference. The OCP classification of unstable and erratic for the files evaluated were as follows: K-Flex (16%), Flex-R (12%), Flex-O (75%), Union Broach NiTi (62%), and Tulsa NiTi (0%). After OCP testing, each of the 120 files was inspected by light microscopy at x 25. The frequencies of visually observed corrosion were detected as follows: K-Flex (2/24), Flex-R (1/24), Flex-O (6/24), Union Broach NiTi (2/24), and Tulsa NiTi (0/24). There was a significant difference in corrosion frequency between brands when evaluated by OCP and light microscopy; however, there was no significant difference between stainless steel and NiTi.
More Related Videos
06:48Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Related Concept Videos
Corrosion
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Microbial Corrosion