Related Experiment Videos
Dilute ornitin vasopressin and a myoma drill for laparoscopic myomectomy
A Rossetti1, M Paccosi, O Sizzi
1Complesso Integrato Columbus, Department of Obstetrics and Gynecology, UCSC Rome, Italy.
Abstract:
To improve enucleation of uterine myomas, we designed a reusable 5-mm laparoscopic drill with a distal forked pin and locking system. Because the device is jointed, it can apply traction not only along its own axis, but also along planes, depending on its bending radius. We compared enucleation time for myomas 5 to 8 cm in size before and after introduction of the drill. The last 23 procedures performed with this instrument were easier, with a reduction in operating time of about 28.5% for the enucleation part of the procedure (p <0.001). (J Am Assoc Gynecol Laparosc 6(2):189-193, 1999)