Electromagnetic interference of an external temporary pacemaker during maxillofacial and neck surgery
1Department of Dental Anesthesia, Niigata University Dental Hospital, Japan.
Anesthesia Progress
|April 1, 1996
Summary
Electrocautery can interfere with temporary pacemakers during neck surgery, causing pacing failure. Occipital placement of the electrocautery ground plate is suggested to prevent this interference in patients with pacemakers.
Area of Science:
- Cardiovascular Medicine
- Surgical Technology
- Biomedical Engineering
Background:
- Patients with pre-existing cardiac conditions like atrioventricular block and bundle-branch block may require temporary pacemakers.
- Electrocautery is a common surgical tool used to control bleeding during procedures.
- Electromagnetic interference (EMI) from electrocautery can potentially affect pacemaker function.
Observation:
- A patient undergoing hemimaxillectomy and radical neck dissection required a temporary transvenous demand pacemaker.
- Despite implementing standard precautions against EMI, the patient experienced pacemaker inhibition during electrocautery use.
- The pacing failure was attributed to current dispersion from the active electrocautery electrode.
Findings:
- Indirect inhibition of an external temporary pacemaker by electrocautery was observed.
- Electrocautery-induced current dispersion poses a risk to temporary pacemaker function.
- Standard EMI precautions were insufficient to prevent pacing failure in this case.
Implications:
- Neck surgery in patients with temporary pacemakers necessitates careful consideration of electrocautery use.
- Occipital placement of the electrocautery ground plate is a potential strategy to mitigate EMI risk.
- This case highlights the importance of monitoring pacemaker function during electrocautery in high-risk patients.