[Use of the DICOM standard in a heterogenous environment. Incompatibility or inoperability?]
Related Experiment Video
Updated: Oct 21, 2026

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
Published on: October 27, 2023
1Klinik für Radiologie, Johannes Gutenberg-Universität Mainz.
Purpose:
The DICOM standard, introduced in 1993, is now established in many of today's image modalities. Experiences with DICOM implementations and requirements for future developments are described.
Methods:
First a brief summary of the development of the DICOM standard is given. Based on general observations, different problems with installation, programming and maintenance are outlined.
Results:
The standard consists of mandatory and optional information models and functionalities to enable the use and development of a broad spectrum of applications. But the available options lead to specific applications not compatible with the basic idea of the standard.
Discussion:
For functional communication distinct DICOM services are necessary. This includes DICOM Basic Worklist, Storage Commitment and Performed Procedure Step. A constructive way to achieve inter-operability through voluntary arrangements between manufactures and users will be shown.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...