Sensitive high-performance liquid chromatographic determination of nifedipine in cat plasma following improved sample
I Nakamura1, M Takahashi, H Izumi
1Department of Hospital Pharmacy, Tohoku University School of Dentistry, Sendai, Japan. i-nakam@mail.cc.tohoku.ac.jp
Abstract:
A simple, easy and accurate reversed-phase high-performance liquid chromatographic method is described for the determination of nifedipine in cat plasma. The procedure involves extraction of nifedipine from plasma using a Sep-Pak C18 cartridge and ultraviolet detection at 350 nm. The present method provides the required reproducibility and sensitivity for the determination of low concentrations of nifedipine without interference from plasma components or photodegradation products. The method was validated over the range 1-50 ng/ml nifedipine. Accuracy and precision were, respectively, 97% or more and 5% or less over the concentration range examined. The minimum quantifiable concentration of nifedipine was found to be 1 ng/ml.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: