Stabilization of a percutaneously implanted port catheter system for hepatic artery chemotherapy infusion
N Shindoh1, Y Ozaki, S Kyogoku
1Department of Radiology, Juntendo University Urayasu Hospital, 2-1-1 Tomioka, Urayasu, Chiba 279-0021, Japan.
Abstract:
A port catheter system for hepatic artery infusion chemotherapy was implanted percutaneously via the left subclavian artery in 41 patients for treatment of unresectable liver metastases. The catheter tip was inserted into the gastroduodenal artery (GDA), the end hole was occluded with a guidewire fragment, and a side-hole for infusion was positioned at the bifurcation of the proper hepatic artery and the GDA. The GDA was embolized with steel coils around the infusion catheter tip via a transfemoral catheter. This procedure is designed to reduce the incidence of hepatic artery occlusion and infusion catheter dislocation.