Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators
W J Groh1, S A Boschee, E D Engelstein
1Krannert Institute of Cardiology, Indiana University, Indianapolis, IN, USA.
Circulation
|July 27, 1999
Summary
Patients with implantable cardioverter-defibrillators (ICDs) can safely walk through electronic article surveillance systems. However, lingering near these systems may lead to inappropriate ICD shocks, posing a risk to patient safety.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Inappropriate shocks from implantable cardioverter-defibrillators (ICDs) have been linked to electronic article surveillance (EAS) systems.
- The specific risks associated with ICDs when patients pass through or linger near EAS systems need further clarification.
Purpose of the Study:
- To assess the safety of ICD function when patients walk through or are exposed to EAS systems.
- To investigate potential interactions between ICDs and EAS systems that could lead to adverse events.
Main Methods:
- 170 subjects with ICDs were evaluated during a brief walk-through and prolonged exposure (2 minutes within 6 inches) to three common EAS systems.
- ICD function was monitored during these exposures, including scenarios with ICD pacing.
Main Results:
- No negative effects on ICD function were observed during a 10- to 15-second walk-through of EAS systems.
- Interactions occurred in 19 out of 169 subjects during prolonged exposure; 7 of these were clinically relevant, with a possibility of inappropriate ICD shocks.
- 12 subjects experienced minor interactions.
Conclusions:
- Walking through EAS systems is safe for patients with ICDs.
- Prolonged exposure or lingering near EAS systems can potentially trigger inappropriate ICD shocks.
More Related Videos
Related Concept Videos
Pulse rhythm
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Instrumentation Amplifier
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Instrument transformers, comprising voltage transformers (VTs) and current transformers (CTs), play crucial roles in power substations by providing isolated replicas of current or voltage for measurement and protection purposes. Voltage transformers reduce the primary voltage to levels suitable for relay operation and measurement, while current transformers scale down the primary current. The primary winding of a current transformer often consists of a single turn, achieved by threading the...
Differential Relays
Differential relays are used to protect generators, buses, and transformers by comparing electrical quantities at different points. When a fault occurs, the difference in current between the two points triggers the relay to operate, opening the circuit breaker. Under normal conditions, the current entering (i1) and leaving (i2) a generator are equal. When a fault occurs, however, these currents become unequal, and the difference current flows in the relay operating coil, causing the relay to...
Holter Monitor: 24-Hour Monitoring
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...