Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter?
1Division of Cardiology, Department of Medicine, National Yang-Ming University, School of Medicine, and Veterans General Hospital-Taipei, Taiwan.
Background:
The prospective, randomized study comparing 4- with 8-mm tip electrodes for radiofrequency linear ablation of typical atrial flutter is not available.
Methods And Results:
A total of 104 consecutive patients with typical atrial flutter were randomly assigned to undergo radiofrequency linear ablation using a 4- (Group I, n=54) or 8-mm tip electrode (Group II, n=50) catheter (temperature-control model, preset 70 degrees C). If complete bidirectional isthmus block could not be achieved after 5 pulses, the ablation catheter was changed to the other type; the maximal radiofrequency pulse number was limited to <10 pulses. Complete or incomplete isthmus conduction block was assessed by activation sequence in a multielectrode Halo catheter during low lateral right atrial and proximal coronary sinus pacing. Before shifting to the other catheter type, the 8-mm electrode catheter achieved higher complete isthmus block rate (92% versus 67%, P<0.05) with fewer pulses (2+/-1 versus 3+/-1, P<0.05), shorter procedure time (24+/-15 versus 31+/-12 minutes, P<0.05), and shorter fluoroscopic time (14+/-10 versus 23+/-15 minutes, P<0.05). After 5 failed ablation pulses, 12 (67%) of 18 patients in group I attained complete isthmus block by using an 8-mm tip catheter, but none of 4 patients in group II achieved complete block by changing to a 4-mm tip catheter.
Conclusions:
The 8-mm tip electrodes are more effective than the standard 4-mm length electrodes in linear ablation for typical atrial flutter. This clinical benefit may be of particular value for some patients with broad and/or thick isthmus.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
08:33Bidirectional Electrical and Optoelectronic Interfaces in Healthy and Ischemic Ex Vivo Rat Hearts
Published on: July 18, 2025
Related Concept Videos
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias