Related Experiment Video
Updated: Oct 19, 2026

Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Cross-sectional presurgical implant imaging using tuned aperture computed tomography (TACT)
H Liang1, D A Tyndall, J B Ludlow
1Department of Diagnostic Sciences and General Dentistry, School of Dentistry, University of North Carolina at Chapel Hill, USA.
Objectives:
To investigate tuned aperture computed tomography (TACT) as an alternative to conventional tomography for cross-sectional imaging of potential implant sites.
Methods:
Twenty potential implant sites in three dry human mandibles were selected. Sixteen TACT basis projections were recorded on film using a linear array of source positions. Films were scanned and 20 TACT and iterative TACT slices were reconstructed using TACT Workbench software. Maximum height and width were measured on one representative slice of each site. Results were compared with the ground truth using ANOVA and Wilcoxon signed rank tests.
Results:
The means of relative difference in maximum height were 1.55 and 1.56% for TACT and iterative TACT respectively (P = 0.8643). The means of relative difference in width were 5.96 and 6.36% for TACT and iterative TACT respectively (P = 0.4845). More than 75% measurements were within +/- 1 mm of ground truth for both modalities.
Conclusions:
TACT may provide an alternative to conventional tomography for dental implant imaging. Further investigations should be performed to explore factors affecting image quality.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT