Severe polyethylene wear in uncemented acetabular cup system components: a report of 5 cases
J Patel1, J E Scott, W J Radford
1Department of Orthopaedics, Chelsea & Westminster Hospital, London, England.
The Journal of Arthroplasty
|September 4, 1999
Abstract:
We report 5 cases of severe failure of the polyethylene liner at 4 to 7.4 years after uncemented anatomic medullary locking (AML) total hip arthroplasty, using the acetabular cup system (ACS). The mode of failure was by a combination of wear-through to the metal backing and liner fracture. Four of the 5 cases were associated with a 32-mm head. The ACS design is flawed because of a lack of hemispherical geometry resulting in rim loading on a region of the liner that has the thinnest polyethylene. All patients with this type of implant should be monitored for evidence of excessive polyethylene wear.