Related Experiment Video
Updated: Aug 15, 2026

Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Clinical validation of an automatic device measuring blood pressure in the fingers
1Department of Vascular Medicine and Rehabilitation, University of Verona. Valeggio sul Mincio, Italy.
Background:
To assess the agreement between a new automatic device (FS-20D) using a cuff-oscillometric method to measure arterial blood pressure (BP) in the fingers and a standard mercury sphygmomanometer.
Methods:
The blood pressure measurements were taken in a sequential order, in a sample of both normotensive subjects (n. 57) and slight to moderate hypertensive patients (n. 28) without vascular complications.
Results:
The mean sphygmomanometer-monitor difference was 0.52 +/- 4.57 mmHg for systolic and 0.25 +/- 4.41 mmHg for diastolic values; the agreement limits were: SBP -8.6 divided by 9.6 mmHg, 95% CI: -0.5 divided by 1.5; DBP: -8.6 divided by 9.1 mmHg, 95% CI: -0.7 divided by 1.2. The grade of agreement between the monitor and the sphygmomanometer was "A" (British Hypertension Society) for both systolic and diastolic values (difference of readings < 5 mmHg: 82%; < 10 mmHg: 97% for systolic blood pressure, 98% of diastolic blood pressure).
Conclusions:
The monitor was proved to be reliable with a good level of precision and accuracy. The FS-20D monitor may be used in self-monitoring of blood pressure of patients with slight to moderate hypertension.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure