[Evaluation of the ambulatory blood pressure monitor SAVE 33 model 2]
R de Gaudemaris1, A White, I Pascal
1Service de médecine Interne et cardiologie, hôpital A-michallon, Grenoble.
Unlabelled:
This work tests a new ambulatory blood pressure device: Save 33 II. The evaluation is conducted according to the phases II and IV of the British Hypertension Society (BHS) protocol: the phases I and III (before and after use) are not performed as this device has received the European quality label 'CE' that guarantees a +/- 3 mmHg accuracy before and after 10,000 simulated cycles. Two trained observers (hearing test and certification done according to the BHS recommendations) have performed 7 sequential BP measurements with conventional and automatic method, at the same arm, among 85 patients. The ranges of mean blood pressure are 78-203 mmHg, for SBP and 53-130 mmHg for DBP.
Results:
The Save 33 II monitor satisfies the grade B of BHS evaluation for both systolic and diastolic pressures. This device also satisfies the Association for the Advancement of Medical Instrumentation (AAMI) protocol: mean difference between methods and standard deviation less than 5 +/- 8 mmHg: -2.7 +/- 6.8 mmHg for SBP and -0.86 +/- 6.6 mmHg for DBP. The in-use assessment is conform to BHS criteria for 20 recordings (less than 20% invalid values during day and night); the 5 other recordings have not enough night values, due to batteries failure (2 cases), sleeping discomfort (2 cases) and a cuff withdraw.
Conclusion:
This device, in the tested configuration, agrees the B grade of BHS and can be recommended for ambulatory blood pressure measurement.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement