Related Experiment Videos
Effect of sterilization on the cutting efficiency of rotary nickel-titanium endodontic files
E Rapisarda1, A Bonaccorso, T R Tripi
1Università Di Catania, Italia.
Objective:
Nickel-titanium instruments that offer greater flexibility than instruments made of other metallic materials have been developed. This study evaluated the behavior in vitro of nickel-titanium rotary instruments under repeated sterilization cycles.
Study Design:
To investigate changes in cutting ability and superficial alterations, 36 ProFile instruments, 18 with a taper of .04 and 18 with a taper of .06, were exposed to different sterilization cycles. Samples were divided into 3 groups of 12 files each. The 12 group A samples were exposed to 14 cycles of sterilization for 30 minutes; the 12 group B samples were exposed to 7 cycles of sterilization for 30 minutes; the 12 group C samples were not sterilized and served as a control group. Chemical compositions of the near surface layers of samples of each group were determined by means of Auger spectroscopy.
Results:
The instruments that underwent the greatest number of sterilizations (group A) showed in-depth distributions of chemical composition that were different from those seen in the control group; this was the result of greater amounts of titanium oxide on the surfaces of the sterilized instruments. The files of group A showed a decrease in cutting efficiency in comparison with those of the control group.
Conclusions:
On the basis of this research on nickel-titanium files, we determined that repeated sterilizations under autoclave alter the superficial structure of such instruments.