Xenon versus ceramics: a comparison of two CT X-ray detector systems
Related Experiment Video
Updated: Oct 18, 2026

Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
C R Becker1, R Bruening, M Schaetzl
1Department of Diagnostic Radiology, Ludwig Maximilians University, Munich, Germany.
Purpose:
The purpose of this work was to compare image quality in phantom and patient CT scans acquired by xenon and ceramic CT detector systems.
Method:
High and low contrast resolution and image noise were determined with a standard CT phantom for both detector systems. In patient CT images, the effect on image noise was measured in anatomical regions of interest in the head, lumbar spine, chest, and abdomen.
Results:
In phantom studies, image noise was significantly lower using ceramic versus xenon detectors. Also, in images of the head and lumbar spine, the signal-to-noise ratio was significantly higher with ceramic than with xenon detectors. In chest scans, ceramic significantly reduced beam-hardening artifacts caused by the thoracic spine. However, in abdominal images, the signal-to-noise ratio was not significantly different between ceramic and xenon detector systems.
Conclusion:
For reduced image noise in CT images of the head, lumbar spine, and chest and high resolution CT, ceramic detector systems appear to be superior to xenon detector systems.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Determination of Crystal Structures
X-ray Imaging
Gas Chromatography: Types of Detectors-II
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...