Related Experiment Video
Updated: Aug 3, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Amsorb: a new carbon dioxide absorbent for use in anesthetic breathing systems
J M Murray1, C W Renfrew, A Bedi
1Department of Anaesthetics and Intensive Care Medicine and the School of Pharmacy, The Queen's University of Belfast, Northern Ireland. jmmurray@compuserve.com
Background:
This article describes a carbon dioxide absorbent for use in anesthesia. The absorbent consists of calcium hydroxide with a compatible humectant, namely, calcium chloride. The absorbent mixture does not contain sodium or potassium hydroxide but includes two setting agents (calcium sulphate and polyvinylpyrrolidine) to improve hardness and porosity.
Methods:
The resultant mixture was formulated and subjected to standardized tests for hardness, porosity, and carbon dioxide absorption. Additionally, the new absorbent was exposed in vitro to sevoflurane, desflurane, isoflurane, and enflurane to determine whether these anesthetics were degraded to either compound A or carbon monoxide. The performance data and inertness of the absorbent were compared with two currently available brands of soda lime: Intersorb (Intersurgical Ltd., Berkshire, United Kingdom) and Dragersorb (Drager, Lubeck, Germany).
Results:
The new carbon dioxide absorbent conformed to United States Pharmacopeia specifications in terms of carbon dioxide absorption, granule hardness, and porosity. When the new material was exposed to sevoflurane (2%) in oxygen at a flow rate of 1 l/min, concentrations of compound A did not increase above those found in the parent drug (1.3-3.3 ppm). In the same experiment, mean +/-SD concentrations of compound A (32.5 +/- 4.5 ppm) were observed when both traditional brands of soda lime were used. After dehydration of the traditional soda limes, immediate exposure to desflurane (60%), enflurane (2%), and isoflurane (2%) produced concentrations of carbon monoxide of 600.0 +/- 10.0 ppm, 580.0 +/- 9.8 ppm, and 620.0 +/-10.1 ppm, respectively. In contrast, concentrations of carbon monoxide were negligible (1-3 ppm) when the anhydrous new absorbent was exposed to the same anesthetics.
Conclusions:
The new material is an effective carbon dioxide absorbent and is chemically unreactive with sevoflurane, enflurane, isoflurane, and desflurane.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
08:00A Synthetic Methodology for Preparing Impregnated and Grafted Amine-Based Silica Composites for Carbon Capture
Published on: September 29, 2023
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Inhalational Anesthetics: Overview
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Acute Respiratory Failure-III
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management