Lupus anticoagulant testing with optical end point automation.
R Luddington1, C Scales, T Baglin
1Department of Haematology, Addenbrookes NHS Trust, Cambridge, United Kingdom. rluddington@yahoo.com
Thrombosis Research
|December 10, 1999
Summary
New lupus anticoagulant screening tests offer improved automation compatibility. The silica clotting time and factor V ratio show promise as sensitive and specific alternatives to the kaolin clotting time.
Area of Science:
- Hematology
- Clinical Chemistry
Background:
- Lupus anticoagulant activity is screened using dilute Russell viper venom time and kaolin clotting time (KCT).
- Kaolin clotting time (KCT) has high turbidity, limiting its use in automated optical detection systems.
- There is a need for alternative lupus anticoagulant screening tests compatible with modern automation.
Purpose of the Study:
- To evaluate five recently reported screening tests as potential alternatives to the kaolin clotting time (KCT) for lupus anticoagulant detection.
- To assess the sensitivity, specificity, and compatibility of new assays with automated systems.
Main Methods:
- Evaluated silica clotting time, Textarin/Ecarin ratio, Taipan venom time, factor V ratio, and a low-turbidity kaolin clotting time.
- Compared the performance of these tests against KCT using lupus-positive and lupus-negative plasma samples.
- Assessed the impact of warfarin and heparin on the tested assays.
- Investigated phospholipid correction and anticardiolipin antibody positivity.
Main Results:
- Silica clotting time demonstrated comparable sensitivity and specificity to KCT, detecting 10/12 KCT-positive samples.
- Silica clotting time allows for confirmatory phospholipid correction but showed false positives in 2/15 warfarinized samples.
- Factor V ratio was unaffected by warfarin or heparin and identified 3 anticardiolipin-positive samples previously screened as lupus-negative.
- A combination of silica clotting time and dilute Russell viper venom time met requirements for sensitivity, phospholipid dependence, and automation compatibility.
Conclusions:
- Silica clotting time and dilute Russell viper venom time are suitable alternatives for lupus anticoagulant screening, offering automation compatibility and phospholipid dependence.
- The factor V ratio serves as a valuable secondary screening tool for anticoagulated patients and anticardiolipin antibody-positive individuals.