Background:
The main purpose of this study was to evaluate the between-instrument variation of the HPLC method for the measurement of total cholesterol (TC), HDL-cholesterol (HDL-C), LDL-cholesterol (LDL-C), VLDL-cholesterol (VLDL-C), chylomicron cholesterol (CM-C), LDL size, and HDL size. Furthermore, the accuracy of the HPLC was assessed for the determination of TC and HDL-C, compared with CDC reference methods.
Methods:
We used four HPLC instruments with different column-load numbers from 250 to 5000. For accuracy assessment of TC and HDL-C, we used the reference methods recommended by the CDC.
Results:
The values measured by the four instruments were highly correlated with each other (mean r = 0.965), and the absolute mean differences were 4-43 mg/L for TC, 4-30 mg/L for HDL-C, 0-48 mg/L for LDL-C, 7-66 mg/L for VLDL-C, 0-7 mg/L for CM-C, 0.1-0.3 nm for LDL size, and 0-0.1 nm for HDL size. For TC, the HPLC instruments showed high correlation and good agreement with the reference method: r = 0.997; total error <6.6%; absolute mean bias <1.2%. For HDL-C, the results from the HPLC method were significantly higher (10.8% absolute mean bias) than those of the CDC reference method, in spite of good correlation between the two methods (r = 0.998).
Conclusions:
The between-instrument variation in serum lipoprotein analysis by HPLC was confirmed to be very small. This method met the US National Cholesterol Education Program's performance criteria for TC but not for HDL-C.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
Blood Studies for Cardiovascular System III: Serum Lipid Profile
Serum lipids are fats and fatty substances in the blood and are crucial for various bodily functions, including energy storage, cellular structure, and hormone production. Serum lipids consist of cholesterol, triglycerides, and phospholipids.
Cholesterol is a soft, fat-like substance found in all body cells. It is crucial for producing hormones, vitamin D, and substances that aid...