Assessment of resting perfusion with myocardial contrast echocardiography: theoretical and practical considerations
J R Lindner1, F S Villanueva, J M Dent
1Cardiovascular Divisions, University of Virginia Medical Center, Charlottesville 22908, USA.
Background:
The aim of this study was to perform a quantitative comparison between myocardial contrast echocardiography (MCE) and single-photon emission computed tomography (SPECT) in patients with prior myocardial infarction (MI). We also wanted to determine the optimal method for the intravenous administration of an ultrasound contrast agent in the clinical setting.
Methods And Results:
Seventeen patients with resting perfusion defects in a single vascular territory on SPECT were studied. MCE was performed with intermittent harmonic imaging during continuous infusions of a second-generation ultrasound contrast agent (Sonovue, Bracco Diagnostics) in all 17 patients and after bolus injection in 8 of them. During continuous infusions, the video intensity (VI) ratio between the abnormal and normal myocardium at a pulsing interval (PI) of 8 cardiac cycles correlated well with the activity ratio between these segments on SPECT (r = 0.73, P <.01). When information regarding microbubble velocity (MV) denoted as change in VI with increasing PIs was added, the correlation with SPECT activity ratio improved (P <.05) significantly (r = 0.87, P <.0001). Higher microbubble doses resulted in higher VI during continuous infusions with good myocardial opacification and no far-field attenuation until the highest dose was reached. With bolus injections, the VI ratio between the abnormal and normal myocardium at PI of 1 and 5 cardiac cycles showed a modest correlation (r = 0.46 and r = 0.48, respectively, P <.05) with activity ratios between these regions on SPECT. When a dose of microbubbles administered as a bolus produced adequate myocardial opacification, it invariably resulted in far-field attenuation.
Conclusions:
In patients with prior MI, quantitative assessment of resting perfusion defects on MCE correlates well with regional activity on SPECT. Continuous infusions offer an advantage over bolus injections because they can provide an assessment of both relative VI and MV. Adjustment of the microbubble infusion rate produces adequate myocardial opacification without attenuation.
More Related Videos
09:10Normothermic Ex Situ Heart Perfusion in Working Mode: Assessment of Cardiac Function and Metabolism
Published on: January 12, 2019
08:21Functional Assessment of the Donor Heart During Ex Situ Perfusion: Insights from Pressure-Volume Loops and Surface Echocardiography
Published on: October 11, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Acute Coronary Syndrome III: Diagnostic Studies