Dual mode antimony electrode for simultaneous measurements of PO2 and pH
1Department of Anesthesiology and Intensive Care, Linköping University Hospital, Sweden.
Background:
In biomedical research and clinical medicine there is a demand for potent sensors to measure the components that make up blood gas analyses. Today, as when the electrochemical PO2, PCO2 and pH electrodes were first introduced, these measurements are usually made with the same type of sensor technology. The aims of the present study were, firstly, to find out whether the platinum cathode in the Clark electrode can be replaced by antimony for oxygen measurements (amperometry (A)); secondly, whether, during oxygen measurements, the inherent corrosion potential of the antimony metal can be used for measurement of pH in the same measurement area (potentiometry (P)).
Methods:
An electrode of purified, crystallographically orientated monocrystalline antimony (COMA) connected to a reference electrode (silver-silver chloride) was used for the P measurements. Measurements of A (at -900 mV) and P were made in an aqueous environment regulated for oxygen, pH, and temperature.
Results:
Reproducible oxygen sensitivities of 0.925 nA/% oxygen (2% CV (coefficient of variation)) (A), 10.7 mV/% (P), and 0.7 mV/% (P) were found in the oxygen range: 0-21%, <5%, and above 5%, respectively. The pH sensitivity was 57 mV/pH unit (P). Oxygen and pH measurements were less accurate at oxygen concentrations close to 0%.
Conclusions:
Both the oxygen and pH part of the composite electrode signal can be identified by this dual mode technique (A and P). The sensor seems to be promising as it provides measurements of two separate variables (oxygen and pH) and also has the desirable characteristics of a solid state sensor.
More Related Videos
13:09Assessment of Boron Doped Diamond Electrode Quality and Application to In Situ Modification of Local pH by Water Electrolysis
Published on: January 6, 2016
08:03Multi-analyte Biochip MAB Based on All-solid-state Ion-selective Electrodes ASSISE for Physiological Research
Published on: April 18, 2013
Related Concept Videos
Potentiometry: Membrane Electrodes
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Amperometry: Overview
Concentration Cells
Consider the following voltaic cell:
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...