Related Experiment Videos
Chromatographic aspects in high throughput liquid chromatography/mass spectrometry
L Pereira1, P Ross, M Woodruff
1ThermoQuest Chromatography Supplies, Hypersil Division, 112 Chadwick Road, Astmoor, Runcorn, Cheshire WA7 1PR, UK.
Abstract:
Important parameters to consider when developing a liquid chromatography/mass spectrometry (LC/MS) method are buffer type and concentration, and column geometry. In the work presented here the choice of buffer for the analysis of basic compounds using a polar embedded phase (HyPURITYtrade mark ADVANCE) is illustrated for the analysis of tricyclic antidepressants. Method transfer from a 4.6 mm i.d. column to a 2.1 mm i.d. column is demonstrated for the analysis of triazines and anabolic steroids and their metabolites, with no change in selectivity and with added speed of analysis. Analysis of eight beta-blockers is achieved in 65 seconds by using a short 30 x 4.6 mm C18 column.