K L Hoffman1, M R Feng, D T Rossi
1Department of Pharmacokinetics and Drug Metabolism, Parke-Davis Pharmaceutical Research, Division of Warner-Lambert Company, Ann Arbor, MI 48105, USA.
Abstract:
A bioanalytical method to quantify cobalt mesoporphyrin (CoMP), a novel therapeutic agent, in plasma has been developed and validated. The approach involves atomic absorption spectroscopy to determine total cobalt in a sample and a back-calculation of the amount of compound present. Endogenous plasma cobalt concentrations were small ( <0.2 ng/ml(-1) Co in rat plasma) in comparison to the quantitation limit (4.5 ng/ml(-1) Co). The inter-day imprecision of the method was 10.0% relative standard deviation (RSD) and the inter-day bias was +/- 8.0% relative error (RE) over a standard curve range of 4.5- 45.0 ng/ml(-1) Co. Because it quantifies total cobalt, the method cannot differentiate between parent drug and metabolites, but negligible metabolism allows reliable estimates of the actual parent drug concentration. A correlation study between the atomic absorption method and 14C-radiometry demonstrated excellent agreement (r = 0.9868, slope = 1.041 +/- 0.028, intercept = 223.7 +/- 190.0) and further substantiated the accuracy of the methods. Methodology was successfully applied to a pharmacokinetic study of CoMP in rat, with pharmacokinetic parameter estimation. The elimination half-lives, after intra-muscular and subcutaneous administration, were 7.7 and 8.8 days, respectively.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Spectroscopy: Absorption, Emission, and Fluorescence