Channelized hotelling and human observer correlation for lesion detection in hepatic SPECT imaging.
H C Gifford1, M A King, D J de Vries
1Department of Radiology, University of Massachusetts Medical School, Worcester 01655, USA.
Summary
Mathematical model observers, like the channelized Hotelling observer (CHO), can predict human performance in medical imaging tasks. This study found strong correlations between human and CHO performance in SPECT liver tumor detection, suggesting CHOs can aid in evaluating imaging strategies.
Area of Science:
- Medical Imaging
- Computational Modeling
- Observer Performance Studies
Background:
- Mathematical model observers are valuable for predicting human performance in medical imaging, potentially replacing human observers in psychophysical studies.
- The channelized Hotelling observer (CHO) is a type of model observer of interest for these applications.
Purpose of the Study:
- To examine the correlations between human observers and several forms of the channelized Hotelling observer (CHO).
- To evaluate the effectiveness of scatter correction strategies in single-photon-emission computed tomography (SPECT) liver imaging for tumor detection.
Main Methods:
- A receiver operating characteristic (ROC) study was conducted using simulated SPECT liver images with varying scatter conditions and correction strategies.
- Human observer performance was assessed by averaging areas under the ROC curve for 7 individuals across different tumor locations and scatter handling strategies.
- Signal-to-noise ratios (SNRs) from human performance were compared with SNRs from constant-Q and difference-of-Gaussian (DOG) CHO models.
Main Results:
- A strong Spearman rank correlation (0.92, P = 0.000001) was observed between average human performance and a constant-Q CHO model.
- A significant correlation (0.84, P = 0.00005) was also found between human performance and the DOG version of the CHO.
Conclusions:
- The significant positive correlations indicate that channelized model observers can potentially replace human observers in certain medical imaging tasks.
- The studied CHO models are effective for preliminary screening of different imaging strategies before conducting human psychophysical studies.