Dose response characteristics and basic dose distribution data for a polymerization-based dosemeter gel evaluated

P Haraldsson1, S A Bäck, P Magnusson

  • 1Department of Radiation Physics, Lund University, Malmö University Hospital, Sweden.

The British Journal of Radiology
|March 18, 2000
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Comparing the Structured Multidisciplinary work Evaluation Tool (SMET) questionnaire with technical measurements of physical workload in certified nursing assistants in a medical ward setting.

Applied ergonomics·2021
Same author

The timed up and go test predicts fracture risk in older women independently of clinical risk factors and bone mineral density.

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA·2020
Same author

An anti-inflammatory phenotype in visceral adipose tissue of old lean mice, augmented by exercise.

Scientific reports·2019
Same author

Anaerobic treatment of oil-contaminated wastewater with methane production using anaerobic moving bed biofilm reactors.

Water research·2019
Same author

Prospective study of growth and bone mass in Swedish children treated with the modified Atkins diet.

European journal of paediatric neurology : EJPN : official journal of the European Paediatric Neurology Society·2019
Same author

Structured Multidisciplinary Work Evaluation Tool (SMET): Reliability testing of a multidisciplinary/multifactorial work questionnaire.

Work (Reading, Mass.)·2019
Same journal

An Evaluation of Radiotherapy and Response in the Management of Perivascular Epithelioid Cell Tumors.

The British journal of radiology·2026
Same journal

Ensuring radiology reporting quality across a national lung cancer screening programme.

The British journal of radiology·2026
Same journal

Utility of High-Resolution Semiconductor Positron Emission Tomography/Computed Tomography in the Assessment of Breast Cancer Extent: Comparison with Magnetic Resonance Imaging.

The British journal of radiology·2026
Same journal

Airway Imaging Practices in Primary Ciliary Dyskinesia: A Global Survey to Guide Standardized Guidelines.

The British journal of radiology·2026
Same journal

Letter to the editor: A comment on statistical power in high-dose radiation group analysis.

The British journal of radiology·2026
Same journal

"Comparative diagnostic accuracy of dual-energy CT, diffusion-weighted MRI and chemical shift MRI in cervical lymph node metastasis in head and neck cancer".

The British journal of radiology·2026
See all related articles

A new method for creating polymerization-based dosemeters using MRI (PoMRI) offers a safe and reproducible way to measure radiation doses. This 3D detector shows excellent potential for accurate dose assessment in radiotherapy.

Area of Science:

  • Medical Physics
  • Radiotherapy
  • Radiation Dosimetry

Background:

  • Accurate radiation dose measurement is critical in radiotherapy.
  • Existing dosimetry methods have limitations in spatial resolution and reproducibility.
  • Polymerization-based MRI (PoMRI) offers a novel approach to 3D dosimetry.

Purpose of the Study:

  • To present a safe and reproducible mixing procedure for PoMRI gel.
  • To evaluate the dose response and sensitivity of PoMRI.
  • To compare PoMRI with diode measurements and assess phantom material suitability.

Main Methods:

  • Developed a mixing procedure for polymerization-based dosemeters.
  • Irradiated gel-filled vials and evaluated dose response using MRI (1/T1 and 1/T2).

Related Experiment Videos

  • Compared PoMRI measurements with diode data for photon and electron beams.
  • Investigated Barex as a phantom material.
  • Main Results:

    • Linear dose response (0-8 Gy) with high sensitivity (0.211 s-1Gy-1 at 1.5 T for 1/T2).
    • PoMRI showed <3% deviation compared to diode measurements for dose curves and profiles.
    • Barex proved a favorable phantom material compared to glass.
    • Background correction methods impact result reliability.

    Conclusions:

    • PoMRI is a safe, reproducible, and accurate 3D radiation dosimetry method.
    • PoMRI demonstrates excellent potential for clinical applications in radiotherapy.
    • Further optimization of background correction and phantom materials can enhance PoMRI accuracy.