Related Experiment Video
Updated: Aug 7, 2026

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Has the mammography quality standards act affected the mammography quality in North Carolina?
E D Pisano1, M Schell, J Rollins
1Department of Radiology, University of North Carolina, Chapel Hill 27599-7510, USA.
Objective:
The United States Food and Drug Administration implemented federal regulations governing mammography under the Mammography Quality Standards Act (MQSA) of 1992. During 1995, its first year in implementation, we examined the impact of the MQSA on the quality of mammography in North Carolina.
Materials And Methods:
All mammography facilities were inspected during 1993-1994, and again in 1995. Both inspections evaluated mean glandular radiation dose, phantom image evaluation, darkroom fog, and developer temperature. Two mammography health specialists employed by the North Carolina Division of Radiation Protection performed all inspections and collected and codified data.
Results:
The percentage of facilities that met quality standards increased from the first inspection to the second inspection. Phantom scores passing rate was 31.6% versus 78.2%; darkroom fog passing rate was 74.3% versus 88.5%; and temperature difference passing rate was 62.4% versus 86.9%.
Conclusion:
In 1995, the first year that the MQSA was in effect, there was a significant improvement in the quality of mammography in North Carolina. This improvement probably resulted from facilities' compliance with federal regulations.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Standards of Care I
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...