Development and characterization of chlorine-selective pulsed discharge emission detector for gas chromatography
Related Experiment Video
Updated: Oct 17, 2026

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
K Sun1, W E Wentworth, S D Stearns
1Department of Chemistry, University of Houston, TX 77204-5641, USA.
Abstract:
A novel chlorine-selective pulsed discharge emission detector (Cl-PDED) for gas chromatography has been developed based on a reaction of krypton with chlorine and a unique design of the detector. A krypton ion produced in the krypton-doped helium pulsed discharge reacts with chlorinated compounds within the pulsed discharge to produce an excited species of KrCl* which emits at 221-222 nm. The reaction has the following advantages in respect to the detection of chlorinated compounds: (1) the reaction is an ion-molecule reaction that is 100-1000 times faster than a reaction of neutrals, which greatly enhances the sensitivity; (2) the KrCl* emission wavelength is far separated from interfering C emissions at 193. and 247.3 nm; (3) the KrCl* emission is transparent to air and can be recorded without a helium purge of the monochromator. The detector itself has been designed to have the following features: (1) the detector has a microvolume of the pulsed discharge region, ca. 0.35 microl, which increases the discharge power density to enhance the sensitivity; (2) this microvolume detector allows the use of a low flow-rate of approximately 5 ml/min, which enhances the sensitivity by the lower dilution of the column effluent; (3) the pulsed discharge is sufficiently narrow to replace the monochromator entrance slit, which gives much greater light gathering power; (4) the discharge electrodes are protected with a helium purge to prevent carbon deposition on the electrodes. This new Cl-PDED is the most sensitive chlorine-selective detector with a minimum detectability of approximately 50 fg Cl/s. The selectivity to carbon is 1000. There are no significant carbon emission lines in the KrCl* emission wavelength region, but the carbon continuum interference (stray light) limits the selectivity. The selectivity could be increased if a double monochromator were used to diminish the stray light. The detector linear range is over three orders of magnitude from 40 fg Cl to approximately 130 pg Cl, and the dynamic range is approximately 4 orders of magnitude. The relative standard deviation of the elemental response to chlorinated compounds is about 5%.
More Related Videos
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...