Impedance audiometric measurements in clinically normal dogs
L K Cole1, M Podell, K W Kwochka
1Department of Veterinary Clinical Sciences, College of Veterinary Medicine, The Ohio State University, Columbus 43210, USA.
American Journal of Veterinary Research
|April 20, 2000
Summary
Impedance audiometry provides reliable measurements in sedated or anesthetized dogs, with no need for sound-attenuated rooms or pre-tympanometry ear flushing. This noninvasive technique aids in diagnosing canine otitis media.
Area of Science:
- Veterinary Medicine
- Audiology
- Animal Physiology
Background:
- Impedance audiometry is a valuable diagnostic tool in veterinary medicine.
- Establishing reference ranges for impedance audiometric values in dogs is crucial for accurate diagnosis.
- Previous studies have not fully evaluated the impact of sedation, anesthesia, and ear flushing on these measurements.
Purpose of the Study:
- To establish reference impedance audiometric values in clinically normal dogs under sedation or anesthesia.
- To assess the influence of ear flushing on tympanometric measurements.
- To determine the necessity of a sound-attenuated room for acoustic reflex testing in dogs.
Main Methods:
- Impedance audiometry, including tympanometry and acoustic reflex testing, was performed on mixed-breed dogs and Beagles under isoflurane or halothane anesthesia, or sedation.
- Tympanometry was conducted before and after ear canal flushing in a subset of dogs.
- Procedures were carried out in both sound-attenuated and non-sound-attenuated environments.
Main Results:
- Ear canal volume and compliance peak were smaller in Beagles compared to mixed-breed dogs, with ear canal volume correlating with body weight.
- No significant differences were observed in tympanometric values before and after ear canal flushing.
- Acoustic reflex testing yielded comparable results regardless of whether it was performed in a sound-attenuated room.
Conclusions:
- Reference ranges for impedance audiometry in sedated or anesthetized dogs were successfully established.
- Acoustic reflex testing in dogs does not require a sound-attenuated environment.
- Ear flushing prior to tympanometry does not alter impedance audiometric results in clinically normal dogs.
- Impedance audiometry is a potentially useful noninvasive method for diagnosing otitis media in dogs.