Summary
Automatic conveyance systems can cause electromagnetic interference (EMI) with medical equipment. Inductive power supply systems pose a risk within 2 meters of rails, exceeding medical equipment immunity standards.
Area of Science:
- Biomedical Engineering
- Electrical Engineering
- Healthcare Technology
Background:
- Automatic conveyance systems are increasingly used in healthcare settings.
- Potential electromagnetic interference (EMI) from these systems with sensitive medical equipment is a concern.
- Existing international standards address electromagnetic immunity for medical devices.
Purpose of the Study:
- To estimate the electromagnetic interference (EMI) risk posed by automatic conveyance systems to electronic medical equipment.
- To measure electric field intensities generated by self-controlled electric truck systems.
- To assess the potential for EMI with specific medical devices like infusion and syringe pumps.
Main Methods:
- Measured electric field intensities from three different self-controlled electric truck systems.
- Evaluated EMI with infusion pumps and syringe pumps placed 1 meter from the system's rail.
- Compared measured electric field intensities against international electromagnetic immunity standards.
Main Results:
- Two systems with non-contact (inductive) power supply mechanisms exhibited maximum electric field intensities at the supplied current frequency.
- The highest measured electric field intensity was 137.0 dB microV/m, exceeding the international immunity standard for medical equipment.
- Systems with contact power supply mechanisms generated significantly lower electric field intensities, well within the immunity standard.
Conclusions:
- Electronic medical equipment may experience EMI if used within 2 meters of automatic conveyance systems utilizing inductive power supply mechanisms.
- EMI is unlikely to occur with systems employing contact power supply mechanisms, even when medical equipment is placed adjacent to the rail.
- Healthcare facilities must consider the type of power supply in automatic conveyance systems to mitigate potential EMI risks to medical devices.