Sonoclot signature during normal pregnancy
1Department of Obstetrics and Gynaecology, Sahlgrenska University Hospital, Göteberg, Sweden. ulla.kjellberg@datadukten.se
Intensive Care Medicine
|April 28, 2000
Summary
This study determined normal ranges for the Sonoclot analyser during pregnancy. Pregnancy is associated with increased hypercoagulability, as indicated by Sonoclot variables, compared to postpartum.
Area of Science:
- Obstetrics and Gynecology
- Hematology
- Medical Device Technology
Background:
- The Sonoclot analyser offers comprehensive hemostasis assessment, including plasma coagulation, platelet function, and fibrinolysis.
- Its utility is established in cardiovascular and hepatic surgery and may aid in managing obstetric complications.
Purpose of the Study:
- To establish reference ranges for Sonoclot analyser variables in normal pregnancy.
- To evaluate changes in hemostasis parameters throughout gestation using the Sonoclot.
Main Methods:
- A prospective, longitudinal study involving 42 healthy pregnant women.
- Sonoclot analysis performed at multiple gestational points (10-15, 32-34, 38-40 weeks) and 8 weeks postpartum.
- Standard laboratory coagulation tests (hemoglobin, hematocrit, platelet count, fibrinogen, APTT) were also analyzed.
Main Results:
- Sonoclot variables indicated a hypercoagulable state during pregnancy, with decreased sonact time and peak time, and increased clot rate and secondary rate compared to postpartum.
- No significant changes in these specific Sonoclot variables were observed within different stages of pregnancy.
- Significant correlations were found between sonact time and activated partial thromboplastin time (APTT), and between clot rate and APTT.
Conclusions:
- The Sonoclot analyser is user-friendly and its signatures are easily interpretable for assessing hemostasis in pregnancy.
- Established Sonoclot ranges are applicable throughout pregnancy, though they differ from postpartum ranges.
- Pregnancy is characterized by significant alterations in hemostasis detectable by the Sonoclot analyser.