Related Experiment Videos
[Carbon dioxide angiography. A new injection system]
A Nicolini1, A Lovaria, D Meregaglia
1Dipartimento di Scienze Radiologiche, Ospedale Maggiore, IRCCS Policlinico, Milano. radzonda@polic.cilea.it
La Radiologia Medica
|May 10, 2000
Summary
The new CO2-Angioset system offers a safe and effective method for digital subtraction angiography, especially for patients with contraindications to iodinated contrast media. This carbon dioxide (CO2) angiography tool demonstrated good visualization in 82% of cases with minimal complications.
Area of Science:
- Vascular Imaging
- Interventional Radiology
- Medical Device Technology
Background:
- Digital subtraction angiography (DSA) traditionally uses iodinated contrast media.
- Iodinated contrast media carry risks, including allergic reactions and nephrotoxicity.
- Alternative contrast agents are needed for specific patient populations.
Purpose of the Study:
- To assess the utility of a novel carbon dioxide (CO2) intravascular injection system (CO2-Angioset) for DSA.
- To evaluate the safety and efficacy of CO2 angiography in various vascular beds.
- To determine the suitability of CO2 angiography for patients with contraindications to iodinated contrast agents.
Main Methods:
- Thirty-nine patients underwent DSA using the CO2-Angioset system between March 1998 and May 1999.
- CO2 angiography was performed on renal arteries, portal vein, and limb vasculature.
- Patients included those with contraindications to iodinated contrast media or poor prior angiography results.
- Hemodynamic and oxygen saturation were monitored; subjective sensations were recorded.
Main Results:
- Adequate visualization was achieved in 82% of CO2 angiography procedures.
- Good depiction of renal arteries and peripheral circulation was noted.
- Portal vein visualization was adequate in 12/18 cases, with limitations in some due to venous anatomy.
- A single transient complication (mesenteric ischemia) occurred, resolving spontaneously.
Conclusions:
- The CO2-Angioset system is a simple, safe, and effective tool for CO2 angiography.
- It is particularly beneficial for patients at risk of allergic reactions to iodinated contrast agents or with renal impairment.
- The system facilitates interventional procedures like arteriovenous fistula embolization and transjugular portosystemic shunting.