An automatic perimeter for glaucoma visual field screening and control. Construction and clinical cases
Summary
This study introduces an automated static perimetry device for assessing the central visual field. The new apparatus offers accurate visual field defect analysis, comparable to conventional methods.
Area of Science:
- Ophthalmology
- Medical instrumentation
- Visual science
Background:
- Static perimetry is crucial for diagnosing visual field defects.
- Conventional perimetry methods can be time-consuming and subjective.
- Automated systems offer potential for improved efficiency and objectivity.
Purpose of the Study:
- To describe a novel apparatus for automated static perimetry.
- To evaluate the performance of this new instrument in analyzing central visual field defects.
Main Methods:
- Development of a fully automatic static perimetry apparatus.
- Programming test logic into a mini-computer for controlled stimulus presentation.
- Plotting results in polar coordinates for clear visualization.
- Analysis of visual field defects using both conventional and automated methods.
Main Results:
- The apparatus successfully performs static perimetry at 64 points in the central visual field.
- Automated testing logic ensures consistent and programmed test execution.
- Results from the automated instrument show satisfying agreement with conventional analysis for small to medium visual field defects.
Conclusions:
- The described automated static perimetry apparatus is a viable tool for visual field analysis.
- This instrument provides accurate and efficient assessment of central visual field defects.
- The automated system demonstrates potential for clinical application in ophthalmology.