Flow-injection analysis of dopamine in injections with a periodate-selective electrode
1CEQUP/Departamento de Química-Física, Faculdade de Farmácia, R. Aníbal Cunha, Porto, Portugal. mcbranco@mail.ff.up.pt
Abstract:
Dopamine determination in pharmaceutical preparations based on its oxidation with periodate (IO(4)(-)) using a new IO(4)(-)-selective electrode under flow conditions is presented. An electrode with a tubular configuration, no internal reference solution, and a PVC (31. 2%) membrane, with metaperiodate bis(triphenylphosphoranylidene)ammonium (1.3%) as ion exchanger and 2-nitrophenyloctylether (67.5%) as mediator solvent, was used. Optimization procedures were directed at potentials versus dopamine readings instead of potential versus the remaining IO(4)(-). This approach was achieved by selecting a 50-cm reactor and an overall flow of 7 mL/min, and injecting 70 microL of dopamine standards in a 3.0 x 10(-4) M IO(4)(-) solution. Under these conditions, a linearity range of 8.0 x 10(-3) to 2.7 x 10(-1) g/L, with a slope of 310.1 +/- 7.4 mV L g(-1) and a reproducibility of +/-0.4 mV, were recorded (n = 8). Interference from common excipients was negligible. Under these conditions, analysis of dopamine injections (n = 12) presenting 200 mg/injection gave average and standard deviation values of 201.0 and 3.3 mg/injection, respectively. A simple and inexpensive flow-injection analysis (FIA) manifold, with a good potentiometric detector, enabled the analysis of 200 samples/h without requiring pretreatment procedures. Comparison with the dopamine injection analysis in the United States Pharmacopoeia monograph showed good accuracy, with a relative deviation of -0.2%.