Related Experiment Videos
Quality control in a manual and an automated leukocyte differential count
1Department of Clinical and Laboratory Medicine Osaka City University Medical School, Osaka, Japan.
Abstract:
Quality control (QC) has been introduced in laboratories, and QC surveys in leukocyte differential count to enhance quality have been performed by College of American Pathologists, Japanese Association of Medical Technologists, Osaka Medical Association and manufacturers. The results of QC survey in a manual leukocyte differential count indicated problems on the differentiation of segmented neutrophils and band neutrophils and the detection of pathological blood cells on blood smear. While the results of QC survey in an automated leukocyte differential count performed by same manufacturer with an automated blood cell counter were satisfactory, however, there was a difference in leukocyte differential cell counts among laboratories with other manufacturer's instruments because the synthetic blood material used in QC is an exclusive item for an instrument. It is necessary to further reeducate the medical technologists in order to improve morphological performance, and to standardize the synthetic blood material for compatibility with various automated blood cell counters.