Measuring Carbon-based Contaminant Mineralization Using Combined CO2 Flux and Radiocarbon Analyses
Published on: October 21, 2016
CO(2) concentration measurements on air samples by mass spectrometry
1Climate and Environmental Physics, Physics Institute, University of Bern, Switzerland.
Abstract:
A new technique for measuring CO(2) concentration in air samples, based on mass spectrometry, is described as an alternative to the common gas chromatographic method. Using a dual inlet isotope ratio mass spectrometer (IRMS), the ratio of the abundances of the m/z peaks 44 and 28 is determined. The precision of measurements (standard deviation <3 ppmv) is generally as good as the analysis with gas chromatography for small air samples (<1 ml STP of air). A major advantage of this new method is the possibility of parallel elemental and isotopic measurements of many air components. The technique is further improved by new wide mass range mass spectrometers allowing simultaneous intensity measurements of several m/z values between 28 and 44, resulting in an uncertainty of <0.5 ppm. The precision is somewhat limited by the production of N(2)O and NO(2) from N(2) and O(2) in the ion source, which accounts for about half of the signal strength at m/z 44. Copyright 2000 John Wiley & Sons, Ltd.
Related Concept Videos
Mass Spectrometry: Overview
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometers
Mass Spectrometry: Isotope Effect
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Measurement of Air Content in Concrete
The pressure method,...